Yokogawa electric production of zirconia oxygen analyzer is widely used in a variety of the combustion monitoring and control process of the industry, and to help industry field has achieved considerable energy saving effect.Application areas including energy consumption industries, such as steel, electric power, petroleum chemical industry, ceramics industry, paper industry, food industry, textile industry, include all sorts of combustion equipment, such as incinerator, middle and small boiler, etc.The analyzer was applied to the combustion monitoring and control, will help full combustion, reducing emissions of CO2, SOx and NOx, and thus contribute to prevent global warming and air pollution.
ZR402G converter with a LCD touch screen, set with a variety of display, calibration, oxygen concentration trend shows that more easy to operate, and shows more enhancements.Converter has all kinds of standard features, such as measuring and computing, as well as the maintenance function, including self-checking.ZR22G points size sensor adopts highly reliable zirconia sensor, the heating equipment can be field replacement.If the sensor is installed in the flue wall, can be directly gas measurement.
features
Points size converter ZR402G with LCD touch screen, easy to operate.
Embedded probe heating device can field replacement, reduced maintenance costs.
Probe with long service life, high reliability of the zirconia sensor.
Using a digital communications (HART) for remote maintenance and reduced maintenance costs.
Description: HART is the HART communication foundation, a registered trademark.
specifications
Measurement object | Combustion exhaust gas and oxygen concentration of mixed gas |
Measurement system | Zirconia system |
Oxygen concentrations | 0.01~100vol% |
The output signal | 4 ~ 20 ma DC (550 Ω maximum load resistance) |
Measuring range | 0 ~ 5 vol % or 0 ~ 100 vol % (O2Unit 1 vol %) range arbitrarily set, or part of the range |
Digital communications (HART) | 250 ~ 550 Ω, depends on the field devices are connected to the circuit number (multipoint mode) |
Sample gas pressure | -5~+250kPa |
The sample temperature | 0 ~ 700 ℃ (probe only) 700 ~ 1400 ℃ (with 0.15 m probe and high temperature probe adapter) |
The length of the probe | 0.15,0.4,0.7,1.0,1.5,2.0,2.5,3.0,3.6,4.2,4.8,5.4m |
Contact output | (fail safe, normally open) at four o 'clock |
Contact input | At two o 'clock |
calibration | Methods: zero/span gas calibration Calibration mode: Automatic, semi-automatic and manual (all through the LCD touch screen interaction) |
The sensor structure | The explosion-proof equates to NEMA4X/IP66 |
The converter structure | Outdoor installation is the same as NEMA4X/IP66 |
The environment temperature | Probe: - 20 ~ 150 ℃ Converter: - 20 ℃ ~ 55 ℃ |
The power supply | 85~264V AC,45~66Hz |
performance
repetitive | Set a maximum range of plus or minus 0.5% |
linear | Set a maximum range of plus or minus 1% |
drift | To set a maximum range of + 2% per month |
The response time | 90% response for 5 seconds |