GC8000 built-in 12 inches color touchscreen, greatly simplifies the maintenance work.As long as the touch screen, all the analysis parameters and the measurement result in the form of simple graphics rendering on the color screen for the technician to check.
GC8000 also in gas chromatograph analysis ability, there are many significant innovation.With GC unit (GCM) concept, for the first time realized the practical application of parallel analysis.Through a single analyzer set in multiple virtual chromatography, make all chromatographic Settings, display and data completely independent, easy to understand and maintain.And, the built-in graphic overview images can be used to check each unit of GC.
GC8000 provides the reliability have been verified and operational aspects of hardware and electronic technology, and through constant temperature furnace analysis ability make the industrial application of gas chromatography has reached a higher level.
GC8000 communications network based on industry standard Ethernet structure, flexible ensures data safely transmitted to a gas chromatograph maintenance station and plant DCS system.GC8000 both Ethernet single configuration, also can use two completely separate Ethernet redundant network.
features
Implementation of the constant temperature furnace structure analysis at a high speed
Large touch screen to improve operating performance
Highly reliable
Performance improvement
specifications
Measurement object | Gas or volatile liquid, the boiling point is 400 ℃ or less) |
Analysis method | Gas chromatography |
Measuring range | Depending on the analysis of conditions TCD: 1ppm~100% FID: 1ppm~100% FID (with methanator) : 1 PPM ~ 0.1% FPD: 1ppm~0.1% |
Measuring group scores | Up to 999 (including all the components of the total flow calibration flow) |
Analysis of the cycle | Maximum of 21600.0 seconds (6 hours) |
Measuring the flow ' | Up to 31 (including calibration flow) |
Contact materials | The RV: 316 ss, hartz alloy C, Rulon, PTFE (teflon, Bearee) LSV: 316 ss, hartz alloy C, Rulon, glass, PTFE (teflon, Bearee), fluorine rubber (Viton), total fluoride rubber (Kalrez) |
repetitive | Depending on the analysis of conditions Gas samples: full range of plus or minus 1% (2 sigma) Liquid samples: full range of plus or minus 2% (2 sigma) |
* according to different specifications and conditions, there will be differences.
Analyzer specification
Protection type | Positive pressure and flame-proof structure, or the explosion-proof structure |
Explosion-proof grade | FM and CSA: Type X Pressurization and Explosionproof for Class I, Division 1,Groups B, C and D. T1~T4 Type X and Y Pressurization for Class I, Division 1,Groups B, C and D. T1~T4 ATEX: II2G Ex d px IIB+H2 T4...T1 Gb IECEx: Ex d px IIB+H2 T1...T4 Gb TIIS: Ex pd IIB+H2 T1~T4 |
structure | NEMA3R, equivalent to IP54 (dustproof drip-proof structure) |
Operating conditions | - 10 ~ 50 ℃, 95% RH or less (no condensation) |
The weight of the | Wall type independent model type 1 is about 100 kg 140 kg2 type about 155 kg 190 kg3 type about 200 kg of about 220 kg |
Constant temperature oven | |
volume | Large temperature furnace: about 45 l Standard temperature furnace: about 31 l |
Temperature setting range | 55 ~ 225 ℃ (set to 1 ℃) |
The temperature control precision | Plus or minus 0.03 ℃ |
Temperature control mode | PID control |
Input/output | |
Analog input/output | Up to 16/32 |
Contact input/output | The largest 32/20 point |
PC communication | Ethernet communication Protocol: TCP/IP, FTP, Modbus CP/IP |
DCS communication | RS-422 Protocol: the MODBUS, Y - Protocol |