One, the introduction:

100 h series portable ultrasonic flowmeter is a universal system, complete functions, handheld time difference method, and a variety of optional features and accessories to meet your all kinds of liquid flow measurement requirements.The small, lightweight, rechargeable battery and universal power supply design can make it more can measure over the ideal choice.Equipped with a coupling agent, built-in data recorder, a pair of 5 m sensor cable, sensor fixed clamp, portable aluminum alloy instrument box or customized high strength protection box,
Second, the measuring principle:
100 h series portable ultrasonic flowmeter by using the theory of measuring the ultrasonic transmission time in the pipeline.Medium (liquid) flow velocity in the pipe, and a lag between the time of ultrasonic propagation along the downstream and upstream of medium with linear relationship.As long as measure the ultrasonic propagation time of downstream and upstream respectively, can be obtained on the basis of the linear relationship at various points along the pipeline route on the velocity of the instantaneous average flow velocity.In this way, the medium flow can be obtained by flow rate, pipe cross-sectional area and the Reynolds number.
When ultrasound beam propagation in the liquid, the flow of the fluid produces small changes will make the travel time, and the propagation time of change is proportional to the liquid flow rate, which can calculate the flow rate of the liquid.
Three, the characteristics of the portable ultrasonic flowmeter:
1, built-in data recorder can record the date, the cumulative flow, signal status, working hours, etc.
Line 2, 4 Chinese characters with display instantaneous flow, flow velocity, cumulative flow, signal state, etc.
3, the standard data interface RS232 for networked detection or export record data.
4, OCT output positive and negative, and the static cumulative frequency pulse signal and the original signal (1-9999 KHZ).
5, non-contact measurement, not interfere with the pipe, no pressure loss.
Six, drive circuit capacitance parameter automatic detection.
Four, the technical characteristics of the portable ultrasonic flowmeter:
1, the measured medium: all kinds of liquid.
2, pipe diameter range: DN10mm ~ 6000 mm.
3, velocity range: 0.01 ~ 30 m/s/s.
4, the accuracy of volume flow/measured value plus or minus 1% depending on the application conditions of the measured values of plus or minus 0.5%.
5, the operating temperature: - 30 ~ 40 ℃.
Section 6, power: built-in rechargeable battery 3 AAA battery adapter built-in Ni - H 100 v - 100 vac.
7, resolution: 0.025 cm/s.
8, reaction time: 1 s.
9, operating time: in the case of a full charge > 14 h.Measurement precision, better than 1.0%.
10, repeatability, better than 0.2%.
11, measuring cycle: 500 ms (2 times per second, 128 groups of data per cycle).
12, working power supply: 90?230 vac (built-in nickel metal hydride rechargeable batteries, after sufficient sustainable work 12 hours).
13, operation: 4 * 4 + 2 sets;Take charge indicator.
14, installation method: topical type installation, simple operation, convenient.
Line 15, display: 4 Chinese characters with display instantaneous flow, flow velocity, cumulative flow, signal state, etc.
16, signal output: the isolation of RS232 (FUJI extension protocol).
Five, the other functions:
1, built-in data recorder can record time and cumulative flow, signal status, working hours, etc.
2, the standard data interface RS232 for networked detection or export record data.
3, since the diagnosis, prompt the current working state is normal.
Six, portable ultrasonic flowmeter configuration:
1, host: 1;
1 to 2, sensors:;
Optional sensors:
Standard of small sensor S type 1 (DN15 ~ 100 mm/applicable temperature 0? 70 ℃)
Standard medium type sensor M1: ((~ 700 mm/applicable temperature 0? 70 ℃)
L standard large sensor type 1: (12-6000 - mm/applicable temperature 0? 70 ℃)
High temperature and small sensors S1H type: (DN15 ~ 100 mm/applicable temperature 0? 160 ℃)
High temperature medium type sensor M1H: ((~ 700 mm/applicable temperature 0? 160 ℃)
Standard sensor S1Z small stent type: (DN15 ~ 100 mm/applicable temperature 0? 70 ℃)
Medium-sized stent type sensor M1Z: (applicable standards (~ 700 mm/applicable temperature 0? 70 ℃)
3, steel chains: 1 set (outer bind type sensor);
4, chargers: 1;
Article 5, data line: 1;
Article 6, special cable: 2 (single 5 meters long);
7, coupling agent: 1 bottle;
8, a tape measure: 1 only;
9, protective box: 1 (aluminum alloy).
Seven, portable ultrasonic flowmeter pipeline:
1 pipe: steel, stainless steel, cast iron, cement pipe, copper, PVC, aluminum, glass reinforced plastics, etc all compact pipeline, the lining is allowed;
Pipe inner 2:15 mm to 6000 mm;
Three straight pipe sensor installation points best satisfy the 10 D, upstream downstream 5 D, 30 D to the pump discharge mean diameter (D);
Eight, measuring medium:
1, type: tap water, sea water, industrial sewage, various oil, acid alkali, alcohol, beer can support such as ultrasonic single homogeneous liquid;
2, temperature: 0 ℃ ~ 160 ℃;
3, turbidity, 10000 PPM or less, and the bubble content is small;
4, velocity: 0 ~ + 30 m/s;
5, flow: positive and reverse bidirectional measurement, and can be measured net flows;
Nine, working environment:
1, the temperature of the host: - 30 ℃ ~ 80 ℃;
2, flow sensor temperature: - 40 ℃ ~ 160 ℃;
3, temperature sensor temperature: selected according to user's requirements;
4, host, humidity: 85% RH;
5, flow sensor, humidity: can work immersed depth of 3 m or less;
6, power supply: AC220V DC8 ~ 36 v or AC7 ~ 30 v.
Ten, the application object:
100 h series portable ultrasonic flowmeter using external clip-on sensor to measure liquid flow.Installation process is very simple, the man-machine interface in Chinese, easier to operate.Especially suitable for traffic balance test and traffic monitoring: drinking water, river water and sea water, cooling water, hot water, industrial wastewater, lubricating oil, diesel oil, fuel oil, chemical liquid, etc.Can be used in the field environment.

Eleven, host, shape size:


Twelve, host wiring diagram:

13, optional sensors:

14, portable ultrasonic flowmeter selection table:

Type no.codeThe name says
100H--------Portable ultrasonic flowmeter
Optional sensorS1
S1 standard type sensor (bound type)

M1
Standard M1 type sensor (bound type)

L1
Standard type L1 sensor (bound type)

S1H
High temperature type sensor S1 (bound type)

M1H
High temperature M1 type sensor (bound type)

S1Z
Standard small stent type S1 (bound type)

M1Z
In the standard type stent M1 (bound type)
Signal cable
X5Standard 5 m x 2
X10Standard 10 m x 2

For example: 100 hs1m1x5
Explanation: the portable ultrasonic flowmeter and the standard standard M1 sensors, equipped with standard S1 type sensors, signal cable 5 m x 2.
15, design selection and ordering information:
When ordering, please specify the name of the measured medium, working pressure, medium temperature, flow rate (velocity) range of pipe sizes and pipe network under test conditions, etc.If you have any questions please call the company technical or sales manager.